  • EEVblog 1663 – BM786 Multimeter REPAIR + InEr Error Investigation

    Repairing a customers Brymen BM786 multimeter, and investigating the cause of the InEr error message on the input jack alert feature.
    Plus a giveaway to a local youngster.

    00:00 – A returned BM786 Multimeter with InEr error message
    03:23 – Split jacks and input alert circuit
    05:19 – In-Circuit probing TIP and Low Ohms mode.
    07:19 – A clue?
    08:05 – A need a new desoldering gun. Removing the input jacks
    10:30 – Bodge testing of the input detection threshold
    12:38 – Not one but TWO failed Megger Insulation meters!
    13:50 – Megger testing the connectors at 500V
    15:19 – Testing a 2nd failed InEr error meter
    17:00 – Cleaning the input jacks
    19:11 – Heating up the input jacks
    20:18 – Getting desperate, removing the input jacks
    21:49 – Getting really desparate, removing the input resistors
    22:38 – Reassembling the original repaired unit
    24:26 – Repaired BM786 GIVEAWAY

    Keysight OLED meter repair: https://www.youtube.com/watch?v=CKdeqqo2diI

    Forum: https://www.eevblog.com/forum/blog/eevblog-1663-bm786-multimeter-repair-iner-error-investigation/